会议论文详细信息
8th International Conference on 3D Radiation Dosimetry
Characterization of a commercially-produced chemically stable Fricke gel dosimeter
Penev, K.I.^1 ; Mequanint, K.^1,2
Department of Chemical and Biochemical Engineering, University of Western Ontario, London
ON
N6A 5B9, Canada^1
Biomedical Engineering Program, University of Western Ontario, London
ON
N6A 5B9, Canada^2
关键词: Commercial productions;    Dose dependences;    Dose response;    Energy dependent;    Fricke gel;    Good correlations;    Pilot scale;    Shelf life;   
Others  :  https://iopscience.iop.org/article/10.1088/1742-6596/573/1/012078/pdf
DOI  :  10.1088/1742-6596/573/1/012078
来源: IOP
PDF
【 摘 要 】

The successful manufacturing of a modified Fricke gel with increased shelf life in a pilot scale (20-liter) commercial production facility is reported. The gels remained chemically stable and usable for over 30 days. Linear dose dependence was observed for cumulative doses up to 30 Gy. Dose-depth studies showed a good correlation between the experimental data and theoretical expectations, but the dose response appeared to be energy-dependent.

【 预 览 】
附件列表
Files Size Format View
Characterization of a commercially-produced chemically stable Fricke gel dosimeter 884KB PDF download
  文献评价指标  
  下载次数:29次 浏览次数:22次