科技报告详细信息
Two Dimensional Power Spectral Density Measurement of X-ray Optics with the Micromap Interferometric Microscope.
Yashchuk, V. V. ; Franck, A. D. ; Irick, S. C. ; Howells, M. R. ; MacDowell, A. A.
Technical Information Center Oak Ridge Tennessee
关键词: X-ray optics;    Anisotropy;    Asymmetry;    Distribution;    Mathematical models;   
RP-ID  :  DE2006862009
学科分类:工程和技术(综合)
美国|英语
来源: National Technical Reports Library
PDF
【 摘 要 】

A procedure and software have been developed to transform the area distribution of the residual surface heights available from the measurement with the Micromap interferometric microscope into a two-dimensional (2D) power spectral density (PSD) distribution of the surface height. The procedure incorporates correction of one of the spectral distortions of the PSD measurement. The distortion appears as a shape difference between the tangential and sagittal PSD spectra deduced from the 2D PSD distribution for an isotropic surface. A detailed investigation of the origin of the anisotropy was performed, and a mathematical model was developed and used to correct the distortion. The correction employs a modulation transfer function (MTF) of the detector deduced analytically based on an experimentally confirmed assumption about the origin of the anisotropy due to the asymmetry of the read-out process of the instrument's CCD camera. The correction function has only one free parameter, the effective width of the gate-shaped apparatus function which is the same for both directions. The value of the parameter, equal to 1.35 pixels, was found while measuring the 2D PSD distribution of the instrument self-noise, independent of spatial frequency. The effectiveness of the developed procedure is demonstrated with a number of PSD measurements with different X-ray optics including mirrors and a grating.

【 预 览 】
附件列表
Files Size Format View
DE2006862009.pdf 762KB PDF download
  文献评价指标  
  下载次数:7次 浏览次数:73次