科技报告详细信息
Development of a Single-Axis Edge Detection System.
Hanshaw, R. A.
Technical Information Center Oak Ridge Tennessee
关键词: Edge detection;    Calibration standards;    Coordinates;    Detection;    Glass;   
RP-ID  :  DE2001751343
学科分类:工程和技术(综合)
美国|英语
来源: National Technical Reports Library
PDF
【 摘 要 】

A SIP (Societe Genevoise Instruments de Physique) Trioptic coordinate measuring machine was modified for calibration of high quality single-axis glass standards to an uncertainty of (+-)0.000020 inch. The modification was accomplished through the addition of a frame grabber board, vision software, a high-resolution camera, stepper motors, a two-axis motor controller, and an HP-IB interface card. An existing temperature system (hygrometer, barometer, laser interferometer system, and optics) was retained as part of the system. An existing Hewlett Packard computer was replaced with a personal computer to accommodate the frame grabber board. Each component was integrated into the existing system using Visual Basic. The system was automated for unattended measurements by creating a machine programming language, which is recognized within the main program.

【 预 览 】
附件列表
Files Size Format View
DE2001751343.pdf 590KB PDF download
  文献评价指标  
  下载次数:14次 浏览次数:25次