The Gas Flow from the Gas Attenuator to the Beam Line
Ryutov, D.D.
SLAC National Accelerator Laboratory
关键词: Light Sources;    Xfel;    43 Particle Accelerators;    Leaks;    Linear Accelerators;   
DOI  :  10.2172/993720
RP-ID  :  SLAC-TN-10-063
RP-ID  :  AC02-76SF00515
RP-ID  :  993720
美国|英语
来源: UNT Digital Library
PDF
【 摘 要 】

The gas leak from the gas attenuator to the main beam line of the Linac Coherent Light Source has been evaluated, with the effect of the Knudsen molecular beam included. It has been found that the gas leak from the gas attenuator of the present design, with nitrogen as a working gas, does not exceed 10{sup -5} torr x l/s even at the highest pressure in the main attenuation cell (20 torr).