| LASER ABLATION-INDUCTIVELY COUPLED PLASMA-ATOMIC EMISSION SPECTROSCOPY STUDY AT THE 222-S LABORATORY USING HOT-CELL GLOVE BOX PROTOTYPE SYSTEM | |
| CM, SEIDEL ; J, JAIN ; JW, OWENS | |
| Hanford Site (HNF), Richland, WA | |
| 关键词: Spectrometers; Testing; Radioactive Waste Facilities; Emission Spectroscopy; Gloveboxes; | |
| DOI : 10.2172/949753 RP-ID : RPP-RPT-39823 Rev 0 RP-ID : DE-AC27-08RV14800 RP-ID : 949753 |
|
| 美国|英语 | |
| 来源: UNT Digital Library | |
PDF
|
|
【 摘 要 】
This report describes the installation, testing, and acceptance of the Waste Treatment and Immobilization Plant (WTP) procured laser ablation-inductively coupled plasma-atomic emission spectroscopy (LA-ICP-AES) system for remotely analyzing high-level waste (HLW) samples in a hot cell environment. The work was completed by the Analytical Process Development (APD) group in accordance with Task Order 2005-003; ATS MP 1027, Management Plan for Waste Treatment Plant Project Work Performed by Analytical Technical Services. The APD group at the 222-S Laboratory demonstrated acceptable turnaround time (TAT) and provide sufficient data to assess sensitivity, accuracy, and precision of the LA-ICP-AES method.
【 预 览 】
| Files | Size | Format | View |
|---|---|---|---|
| 949753.pdf | 2153KB |
PDF