科技报告详细信息
Precision volume measurement system.
Fischer, Erin E. ; Shugard, Andrew D.
Sandia National Laboratories
关键词: Measuring Instruments;    Pressure Measurement;    47 Other Instrumentation;    37 Inorganic, Organic, Physical And Analytical Chemistry;    Measuring Instruments-Reliability.;   
DOI  :  10.2172/921130
RP-ID  :  SAND2004-8105
RP-ID  :  AC04-94AL85000
RP-ID  :  921130
美国|英语
来源: UNT Digital Library
PDF
【 摘 要 】

A new precision volume measurement system based on a Kansas City Plant (KCP) design was built to support the volume measurement needs of the Gas Transfer Systems (GTS) department at Sandia National Labs (SNL) in California. An engineering study was undertaken to verify or refute KCP's claims of 0.5% accuracy. The study assesses the accuracy and precision of the system. The system uses the ideal gas law and precise pressure measurements (of low-pressure helium) in a temperature and computer controlled environment to ratio a known volume to an unknown volume.

【 预 览 】
附件列表
Files Size Format View
921130.pdf 9571KB PDF download
  文献评价指标  
  下载次数:10次 浏览次数:24次