科技报告详细信息
| Calibration of Small Plastic Scintillators for Imaging Applications | |
| Pozzi, S. | |
| Oak Ridge National Laboratory | |
| 关键词: Calibration; Plastic Scintillation Detectors; 46 Instrumentation Related To Nuclear Science And Technology; Phosphors; Plastic Scintillators; | |
| DOI : 10.2172/885841 RP-ID : ORNL/TM-2004/298 RP-ID : DE-AC05-00OR22725 RP-ID : 885841 |
|
| 美国|英语 | |
| 来源: UNT Digital Library | |
PDF
|
|
【 摘 要 】
This report presents the results of measurements and simulations performed with 12 small plastic scintillation detectors manufactured by Scionix for imaging applications. The scintillator is equivalent to a Bicron BC-420 plastic scintillator. A gamma calibration is presented to determine the voltage to be applied on each detector to ensure uniform detector operation. Time of flight measurements performed with a Cf-252 source are also presented. Comparisons between experimental data and data from the Monte Carlo simulations show good agreement for time lags of 0 to 70 ns.
【 预 览 】
| Files | Size | Format | View |
|---|---|---|---|
| 885841.pdf | 197KB |
PDF