【 摘 要 】
This report presents the results of measurements and simulations performed with 12 small plastic scintillation detectors manufactured by Scionix for imaging applications. The scintillator is equivalent to a Bicron BC-420 plastic scintillator. A gamma calibration is presented to determine the voltage to be applied on each detector to ensure uniform detector operation. Time of flight measurements performed with a Cf-252 source are also presented. Comparisons between experimental data and data from the Monte Carlo simulations show good agreement for time lags of 0 to 70 ns.