科技报告详细信息
Calibration of the Accuscan II In Vivo System for I-131 Thyroid Counting
Perry, Orval R. ; Georgeson, David L.
Idaho National Laboratory
关键词: Calibration;    Testing;    Iodine;    Verification Bioassay;    Bioassay;   
DOI  :  10.2172/1042395
RP-ID  :  INL/EXT-11-22507
RP-ID  :  DE-AC07-05ID14517
RP-ID  :  1042395
美国|英语
来源: UNT Digital Library
PDF
【 摘 要 】

This report describes the March 2011 calibration of the Accuscan II HpGe In Vivo system for I-131 thyroid counting. The source used for the calibration was an Analytics mixed gamma source 82834-121 distributed in an epoxy matrix in a Wheaton Liquid Scintillation Vial with energies from 88.0 keV to 1836.1 keV. The center of the detectors was position 64-feet from the vault floor. This position places the approximate center line of the detectors at the center line of the source in the thyroid tube. The calibration was performed using an RMC II phantom (Appendix J). Validation testing was performed using a Ba-133 source and an ANSI N44.3 Phantom (Appendix I). This report includes an overview introduction and records for the energy/FWHM and efficiency calibrations including verification counting. The Accuscan II system was successfully calibrated for counting the thyroid for I-131 and verified in accordance with ANSI/HPS N13.30-1996 criteria.

【 预 览 】
附件列表
Files Size Format View
1042395.pdf 1639KB PDF download
  文献评价指标  
  下载次数:22次 浏览次数:60次