Calibration of the Accuscan II In Vivo System for I-131 Thyroid Counting | |
Orval R. Perry ; David L. Georgeson | |
关键词: CALIBRATION; EFFICIENCY; GAMMA SOURCES; IN VIVO; PHANTOMS; SCINTILLATIONS; TESTING; THYROID; VALIDATION; VERIFICATION Bioassay; Calibration; Iodine; Thyroid; Whole Body Counting; | |
DOI : 10.2172/1042395 RP-ID : INL/EXT-11-22507 PID : OSTI ID: 1042395 Others : TRN: US1202869 |
|
学科分类:核能源与工程 | |
美国|英语 | |
来源: SciTech Connect | |
【 摘 要 】
This report describes the March 2011 calibration of the Accuscan II HpGe In Vivo system for I-131 thyroid counting. The source used for the calibration was an Analytics mixed gamma source 82834-121 distributed in an epoxy matrix in a Wheaton Liquid Scintillation Vial with energies from 88.0 keV to 1836.1 keV. The center of the detectors was position 64-feet from the vault floor. This position places the approximate center line of the detectors at the center line of the source in the thyroid tube. The calibration was performed using an RMC II phantom (Appendix J). Validation testing was performed using a Ba-133 source and an ANSI N44.3 Phantom (Appendix I). This report includes an overview introduction and records for the energy/FWHM and efficiency calibrations including verification counting. The Accuscan II system was successfully calibrated for counting the thyroid for I-131 and verified in accordance with ANSI/HPS N13.30-1996 criteria.
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
RO201704210001876LZ | 1639KB | download |