Chilled Mirror Dew Point Hygrometer (CM) Handbook
Ritsche, MT
Pacific Northwest National Laboratory (U.S.)
关键词: Storage;    Filtration;    Air;    Dew Point;    Humidity;   
DOI  :  10.2172/1020567
RP-ID  :  DOE/SC-ARM/TR-032
RP-ID  :  DE-AC05-7601830
RP-ID  :  1020567
美国|英语
来源: UNT Digital Library
PDF
【 摘 要 】

The CM systems have been developed for the ARM Program to act as a moisture standard traceable to National Institute of Standards and Technology (NIST). There are three CM systems that are each fully portable, self-contained, and require only 110 V AC power. The systems include a CM sensor, air sampling and filtration system, a secondary reference (Rotronic HP043 temperature and relative humidity sensor) to detect system malfunctions, a data acquisition system, and data storage for more than one month of 1-minute data. The CM sensor directly measures dew point temperature at 1 m, air temperature at 2 m, and relative humidity at 2 m. These measurements are intended to represent self-standing data streams that can be used independently or in combinations.