科技报告详细信息
Fluorescence measurements for evaluating the application of multivariate analysis techniques to optically thick environments.
Reichardt, Thomas A. ; Timlin, Jerilyn Ann ; Jones, Howland D. T. ; Sickafoose, Shane M. ; Schmitt, Randal L.
Sandia National Laboratories
关键词: Multivariate Analysis;    Spectra;    Fluorescence;    37 Inorganic, Organic, Physical And Analytical Chemistry;    Laser Radiation;   
DOI  :  10.2172/1008131
RP-ID  :  SAND2010-6571
RP-ID  :  AC04-94AL85000
RP-ID  :  1008131
美国|英语
来源: UNT Digital Library
PDF
【 摘 要 】
Laser-induced fluorescence measurements of cuvette-contained laser dye mixtures are made for evaluation of multivariate analysis techniques to optically thick environments. Nine mixtures of Coumarin 500 and Rhodamine 610 are analyzed, as well as the pure dyes. For each sample, the cuvette is positioned on a two-axis translation stage to allow the interrogation at different spatial locations, allowing the examination of both primary (absorption of the laser light) and secondary (absorption of the fluorescence) inner filter effects. In addition to these expected inner filter effects, we find evidence that a portion of the absorbed fluorescence is re-emitted. A total of 688 spectra are acquired for the evaluation of multivariate analysis approaches to account for nonlinear effects.
【 预 览 】
附件列表
Files Size Format View
1008131.pdf 1518KB PDF download
  文献评价指标  
  下载次数:9次 浏览次数:22次