Majorana Electroformed Copper Mechanical Analysis | |
Overman, Nicole R. ; Overman, Cory T. ; Kafentzis, Tyler A. ; Edwards, Danny J. ; Hoppe, Eric W. | |
关键词: COPPER; DECAY; DESIGN; EVALUATION; GE SEMICONDUCTOR DETECTORS; HARDNESS; METALLOGRAPHY; SCANNING ELECTRON MICROSCOPY; SHIELDING; SPECIFICATIONS; TESTING Majorana; Electroformed; Copper; Mechanical; Properties; | |
DOI : 10.2172/1039850 RP-ID : PNNL-21315 PID : OSTI ID: 1039850 Others : Other: 830403000 Others : TRN: US1202278 |
|
美国|英语 | |
来源: SciTech Connect | |
![]() |
【 摘 要 】
The MAJORANA DEMONSTRATOR is a large array of ultra-low background high-purity germanium detectors, enriched in 76Ge, designed to search for zero-neutrino double-beta decay. The DEMONSTRATOR will utilize ultra high purity electroformed copper for a variety of detector components and shielding. A preliminary mechanical evaluation was performed on the Majorana prototype electroformed copper material. Several samples were removed from a variety of positions on the mandrel. Tensile testing, optical metallography, scanning electron microscopy, and hardness testing were conducted to evaluate mechanical response. Analyses carried out on the Majorana prototype copper to this point show consistent mechanical response from a variety of test locations. Evaluation shows the copper meets or exceeds the design specifications.
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
RO201704190003329LZ | 2027KB | ![]() |