科技报告详细信息
Characterization of the NPOD3 Detectors in MCNP5 and MCNP6
Clark, Kimberly L.1  Hutchinson, Jesson D.1  Sood, Avneet1 
[1]Los Alamos National Lab. (LANL), Los Alamos, NM (United States)
关键词: NEUTRON DETECTORS;    PLUTONIUM-ALPHA;    COMPARATIVE EVALUATIONS;    POLYETHYLENES;    COUNTING RATES;    ABSORPTION;    M CIDES;    COMPUTERIZED SIMULATION;    NEUTRON LEAKAGE;    MULTIPLICATION FACTORS;    THICKNESS;    CONFIGURATION;    EFFICIENCY;    SPHERES;   
DOI  :  10.2172/1115551
RP-ID  :  LA-UR--14-20342
PID  :  OSTI ID: 1115551
Others  :  TRN: US1500365
美国|英语
来源: SciTech Connect
PDF
【 摘 要 】
Researchers performed a series of measurements in May 2012 to characterize the NPOD3 detector systems. The detectors were placed in varying states of disassembly to determine the effect of individual components on detection efficiency. A 4.5 kg ??-phase Pu sphere known as the Los Alamos BeRP Ball was used as the SNM source in both a bare configuration and reflected by varying thicknesses of polyethylene. A set of simulations matching the experimental setups were run and the data were compared to the measured data. The total and leakage multiplication and the inferred k values were determined for both the simulations and the measurements. Table 3 shows a comparison of the results from MCNP6 and MCNP5 with the list-mode patch to the measured results. The count rates for the calculated results were obtained by dividing the total line count in the list-mode file (equivalent to the total number of absorptions in the NPOD detectors) by the total run time. The count rates are identical for both codes, and they both produce the same multiplicity and inferred k values regardless of measurement time as expected.
【 预 览 】
附件列表
Files Size Format View
374KB PDF download
  文献评价指标  
  下载次数:10次 浏览次数:44次