High-Efficiency CdZnTe Position-Sensitive VFG Gamma-Ray Detectors for Safeguards Applications | |
Bolotnikov, Aleksey E.1  James, Ralph B.1  Cui, Y.1  De Geronimo, G.1  Vernon, E.1  Camarda, G. S.1  Hossain, A.1  Yang, G.1  Indusi, J.1  Boyer, Brian2  | |
[1] Brookhaven National Lab. (BNL), Upton, NY (United States);Los Alamos National Lab. (LANL), Los Alamos, NM (United States) | |
关键词: CdZnTe; Position-Sensitive; VFG Gamma-Ray Detectors; Safeguards Applications; | |
DOI : 10.2172/1328361 RP-ID : BNL--112190-2015 PID : OSTI ID: 1328361 |
|
学科分类:核物理和高能物理 | |
美国|英语 | |
来源: SciTech Connect | |
【 摘 要 】
The goal of this project is to incorporate a Cadmium-Zinc-Telluride (CdZnTe or CZT) detector (with 1% or better resolution) into a bench-top prototype for isotope identification and related safeguards applications. The bench-top system is based on a 2x2 array of 6x6x20 mm3 position-sensitive virtual Frisch-grid (VFG) CZT detectors. The key features of the array are that it allows for the use of average-grade CZT material with a moderate content of defects, and yet it provides high-energy resolution, 1% FWHM at 662 keV, large effective area, and low-power consumption. The development of this type of 3D detector and new instruments incorporating them is motivated by the high cost and low availability of large, > 1 cm3, CZT crystals suitable for making multi-pixel detectors with acceptable energy resolution and efficiency.