Determination of lead in human saliva by combined cloud point extraction-capillary zone electrophoresis with indirect UV detection
Article
Luconi, MO ; Olsina, RA ; Fernández, LP ; Silva, MF
关键词: cloud point extraction;    saliva biomarkers;    lead;    capillary zone electrophoresis;   
DOI  :  10.1016/j.jhazmat.2005.08.007
来源: Elsevier
PDF
【 摘 要 】

A micelle-mediated phase separation without added chelating agents to preconcentrate trace levels of lead in human saliva as a prior step to its determination by capillary electrophoresis has been developed. The enrichment step is based on the cloud point extraction of lead with the non-ionic surfactant PONPE 7.5 in the absence of chelating agent. The surfactant-rich phase was diluted with acetonitrile and the resultant solution was injected directly into the CE instrument. Factors affecting the combined methodology such as surfactant-rich phase diluting agent, buffer pH and concentration, applied voltage, sample preparation and presence of additives were studied in detail. A BGE of 20 mM imidazole containing 30% acetonitrile, pH 6.20 was found to be optimal for the separation of lead from other saliva constituents. Indirect detection was performed at 205 nm. The detection limit value of lead for the preconcentration of 8 ml of saliva was 11.4 mu g l(-1). The calibration graph using the preconcentration system was linear with a correlation coefficient of 0.997 at levels near the detection limits up to at least 400 mu g l(-1). The reproducibility (R.S.D.) on the basis of migration time and peak area were better than 0.68 and 3.6%, respectively. The method was successfully applied to the determination of lead in human saliva. (c) 2005 Elsevier B.V. All rights reserved.

【 授权许可】

Free   

【 预 览 】
附件列表
Files Size Format View
10_1016_j_jhazmat_2005_08_007.pdf 107KB PDF download
  文献评价指标  
  下载次数:0次 浏览次数:0次