期刊论文详细信息
TALANTA 卷:94
Development of an automated on-line electrochemical chlorite ion sensor
Article
Myers, John N.2  Steinecker, William H.2  Cox, James A.2  Gordon, Gilbert2  Pacey, Gilbert E.1 
[1] Univ Dayton, Res Inst, Inst Dev & Commercializat Adv Sensor Technol, Dayton, OH 45469 USA
[2] Miami Univ, Dept Chem & Biochem, Oxford, OH 45056 USA
关键词: Chlorite ion;    Electrochemical probe;    Flow injection analysis;   
DOI  :  10.1016/j.talanta.2012.03.026
来源: Elsevier
PDF
【 摘 要 】

A sensor system for the automatic, in-line, determination of chlorite ion is reported. Electroanalytical measurements were performed in electrolyte-free liquids by using an electrochemical probe (EC), which enables in-line detection in high-resistance media such as disinfected water. Cyclic voltammetry scan rate studies suggest that the current arising from the oxidation of chlorite ion at an EC probe is mass-transfer limited. By coupling FIA with an EC probe amperometric cell, automated analysis was achieved. This sensor is intended to fulfill the daily monitoring requirements of the EPA DBP regulations for chlorite ion. Detection limits of 0.02-0.13 mg/L were attained, which is about one order of magnitude below the MRDL. The sensor showed no faradaic signal for perchlorate, chlorate, or nitrate. The lifetime and stability of the sensor were investigated by measuring calibration curves over time under constant-flow conditions. Detection limits of <0.1 mg/L were repeatedly achieved over a period of three weeks. (C) 2012 Elsevier B.V. All rights reserved.

【 授权许可】

Free   

【 预 览 】
附件列表
Files Size Format View
10_1016_j_talanta_2012_03_026.pdf 410KB PDF download