期刊论文详细信息
TALANTA 卷:114
Detection of biothiols in human serum by QDs based flow injection turn off-on chemiluminescence analysis system
Article
Liu, Linlin1  Ma, Qiang1  Li, Yang1  Liu, Ziping1  Su, Xingguang1 
[1] Jilin Univ, Dept Analyt Chem, Coll Chem, Changchun 130012, Peoples R China
关键词: Chemiluminescence;    Flow injection;    Quantum dots;    Glutathione;    Turn off-on;   
DOI  :  10.1016/j.talanta.2013.05.010
来源: Elsevier
PDF
【 摘 要 】

In the present work, a flow injection (FI) turn off-on chemiluminescent method was developed for the determination of glutathione (GSH). Strong chemiluminescence (CL) signals were observed from the hydrogen peroxide and CdTe quantum dots (QDs) system under basic condition, addition of trace amount of Cu(II) could caused significant CL quenching of the CdTe QDs-H2O2 system. In the presence of biothiols, Cu(II) can be removed from CdTe QDs surface via forming Cu(II)-S bond with thiols, and the CL signal of CdTe QD5-H2O2 system was recovered. Thus, the CL signals of CdTe QDs-H2O2 system were turned off and on by the addition of Cu(II) and biothiols respectively, and a flow injection CL analysis system for the determination of biothiols was established. Under the optimum conditions, the CL intensity and the concentration of GSH have a good linear relationship in the range of 2.0 x 10(-9)-6.5 x 10(-7) mol L-1 (R-2=0.9993). The limit of detection for GSH is 1.5 x 10(-9) mol L-1 (S/N=3). This method has been applied to detect GSH in human serum with satisfactory results. (C) 2013 The Authors. Published by Elsevier B.V. All rights reserved.

【 授权许可】

Free   

【 预 览 】
附件列表
Files Size Format View
10_1016_j_talanta_2013_05_010.pdf 614KB PDF download
  文献评价指标  
  下载次数:0次 浏览次数:1次