| The diagnostic accuracy of the MyDiagnostick to detect atrial fibrillation in primary care | |
| Research Article | |
| Wilfried Mullens1  Daphne Lapeire2  Karen Tavernier2  Silke Stalpaert2  Britt Thaels2  Bert Vaes3  Jan Degryse3  | |
| [1] Department of Cardiology, Ziekenhuis Oost Limburg, Genk, Belgium;Biomedical Research Institute, Faculty of Medicine and Life Sciences, Hasselt University, Diepenbeek, Belgium;Department of General Practice, Katholieke Universiteit Leuven (KUL), Leuven, Belgium;Department of General Practice, Katholieke Universiteit Leuven (KUL), Leuven, Belgium;Institute of Health and Society, Université Catholique de Louvain (UCL), Brussels, Belgium; | |
| 关键词: Atrial fibrillation; Sensitivity and specificity; MyDiagnostick; Primary care; | |
| DOI : 10.1186/1471-2296-15-113 | |
| received in 2014-02-01, accepted in 2014-06-05, 发布年份 2014 | |
| 来源: Springer | |
PDF
|
|
【 摘 要 】
BackgroundAtrial fibrillation is very common in people aged 65 or older. This condition increases the risk of death, congestive heart failure and thromboembolic conditions. Many patients with atrial fibrillation are asymptomatic and a cerebrovascular accident (CVA) is often the first clinical presentation. Guidelines concerning the prevention of CVA recommend monitoring the heart rate in patients aged 65 or older. Recently, the MyDiagnostick (Applied Biomedical Systems BV, Maastricht, The Netherlands) was introduced as a new screening tool which might serve as an alternative for the less accurate pulse palpation. This study was designed to explore the diagnostic accuracy of the MyDiagnostick for the detection of atrial fibrillation.MethodsA phase II diagnostic accuracy study in a convenience sample of 191 subjects recruited in primary care. The majority of participants were patients with a known history of atrial fibrillation (n = 161). Readings of the MyDiagnostick were compared with electrocardiographic recordings. Sensitivity and specificity and their 95% confidence interval were calculated using 2x2 tables.ResultsA prevalence of 54% for an atrial fibrillation rhythm was found in the study population at the moment of the study. A combination of three measurements with the MyDiagnostick for each patient showed a sensitivity of 94% (95% CI 87 – 98) and a specificity of 93% (95% CI 85 – 97).ConclusionThe MyDiagnostick is an easy-to-use device that showed a good diagnostic accuracy with a high sensitivity and specificity for atrial fibrillation in a convenience sample in primary care. Future research is needed to determine the place of the MyDiagnostick in possible screening or case-finding strategies for atrial fibrillation.
【 授权许可】
CC BY
© Vaes et al.; licensee BioMed Central Ltd. 2014
【 预 览 】
| Files | Size | Format | View |
|---|---|---|---|
| RO202311108399697ZK.pdf | 470KB |
【 参考文献 】
- [1]
- [2]
- [3]
- [4]
- [5]
- [6]
- [7]
- [8]
- [9]
- [10]
- [11]
- [12]
- [13]
- [14]
- [15]
- [16]
- [17]
- [18]
- [19]
- [20]
PDF