Journal of Cardiovascular Magnetic Resonance
Modified look-locker inversion recovery T1 mapping indices: assessment of accuracy and reproducibility between magnetic resonance scanners
Research
Joao AC Lima1  Melanie Freed2  Chia-Ying Liu3  Fabio S Raman3  Neville Gai3  Nadine Kawel-Boehm4  Songtao Liu5  David A Bluemke5  Jing Han6 
[1] Division of Cardiology, Johns Hopkins University School of Medicine, Baltimore, MD, USA;New York University School of Medicine, New York, NY, USA;Radiology and Imaging Sciences, National Institutes of Health Clinical Center, 10 Center Drive, MSC-1182, 20892-1182, Bethesda, MD, USA;Radiology and Imaging Sciences, National Institutes of Health Clinical Center, 10 Center Drive, MSC-1182, 20892-1182, Bethesda, MD, USA;Clinic of Radiology and Nuclear Medicine, University of Basel Hospital, Basel, Switzerland;Radiology and Imaging Sciences, National Institutes of Health Clinical Center, 10 Center Drive, MSC-1182, 20892-1182, Bethesda, MD, USA;Molecular Biomedical Imaging Laboratory, National Institute of Biomedical Imaging and Bioengineering, Bethesda, MD, USA;U.S. Food and Drug Administration, Rockville, MD, USA;
关键词: T1 mapping;    Partition coefficient (λ);    Extracellular volume fraction (ECV);    Diffuse myocardial fibrosis;    Modified look-locker with inversion recovery (MOLLI);   
DOI  :  10.1186/1532-429X-15-64
 received in 2013-02-24, accepted in 2013-07-19,  发布年份 2013
来源: Springer
PDF
【 摘 要 】

BackgroundCardiovascular magnetic resonance (CMR) T1 mapping indices, such as T1 time and partition coefficient (λ), have shown potential to assess diffuse myocardial fibrosis. The purpose of this study was to investigate how scanner and field strength variation affect the accuracy and precision/reproducibility of T1 mapping indices.MethodsCMR studies were performed on two 1.5T and three 3T scanners. Eight phantoms were made to mimic the T1/T2 of pre- and post-contrast myocardium and blood at 1.5T and 3T. T1 mapping using MOLLI was performed with simulated heart rate of 40-100 bpm. Inversion recovery spin echo (IR-SE) was the reference standard for T1 determination. Accuracy was defined as the percent error between MOLLI and IR-SE, and scan/re-scan reproducibility was defined as the relative percent mean difference between repeat MOLLI scans. Partition coefficient was estimated by ΔR1myocardium phantom/ΔR1blood phantom. Generalized linear mixed model was used to compare the accuracy and precision/reproducibility of T1 and λ across field strength, scanners, and protocols.ResultsField strength significantly affected MOLLI T1 accuracy (6.3% error for 1.5T vs. 10.8% error for 3T, p<0.001) but not λ accuracy (8.8% error for 1.5T vs. 8.0% error for 3T, p=0.11). Partition coefficients of MOLLI were not different between two 1.5T scanners (47.2% vs. 47.9%, p=0.13), and showed only slight variation across three 3T scanners (49.2% vs. 49.8% vs. 49.9%, p=0.016). Partition coefficient also had significantly lower percent error for precision (better scan/re-scan reproducibility) than measurement of individual T1 values (3.6% for λ vs. 4.3%-4.8% for T1 values, approximately, for pre/post blood and myocardium values).ConclusionBased on phantom studies, T1 errors using MOLLI ranged from 6-14% across various MR scanners while errors for partition coefficient were less (6-10%). Compared with absolute T1 times, partition coefficient showed less variability across platforms and field strengths as well as higher precision.

【 授权许可】

Unknown   
© Raman et al.; licensee BioMed Central Ltd. 2013. This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

【 预 览 】
附件列表
Files Size Format View
RO202311104005359ZK.pdf 858KB PDF download
【 参考文献 】
  • [1]
  • [2]
  • [3]
  • [4]
  • [5]
  • [6]
  • [7]
  • [8]
  • [9]
  • [10]
  • [11]
  • [12]
  • [13]
  • [14]
  • [15]
  • [16]
  • [17]
  • [18]
  • [19]
  • [20]
  • [21]
  • [22]
  • [23]
  • [24]
  • [25]
  • [26]
  • [27]
  • [28]
  • [29]
  • [30]
  • [31]
  • [32]
  • [33]
  • [34]
  • [35]
  • [36]
  • [37]
  • [38]
  文献评价指标  
  下载次数:0次 浏览次数:0次