Microbial Cell Factories
Online monitoring of dissolved oxygen tension in microtiter plates based on infrared fluorescent oxygen-sensitive nanoparticles
Research
Tobias Ladner1  David Flitsch1  Jochen Büchs1  Tino Schlepütz1 
[1] AVT, Aachener Verfahrenstechnik, Biochemical Engineering, RWTH Aachen University, Worringerweg 1, 52074, Aachen, Germany;
关键词: High-throughput screening;    Online monitoring;    Dissolved oxygen tension;    Oxygen-sensitive nanoparticles;    Optical measurement;    Microtiter plate;    Fluorescence tags;   
DOI  :  10.1186/s12934-015-0347-9
 received in 2015-07-30, accepted in 2015-09-25,  发布年份 2015
来源: Springer
PDF
【 摘 要 】

BackgroundDuring the past years, new high-throughput screening systems with capabilities of online monitoring turned out to be powerful tools for the characterization of microbial cell cultures. These systems are often easy to use, offer economic advantages compared to larger systems and allow to determine many important process parameters within short time. Fluorescent protein tags tremendously simplified the tracking and observation of cellular activity in vivo. Unfortunately, interferences between established fluorescence based dissolved oxygen tension (DOT) measurement techniques and fluorescence-based protein tags appeared. Therefore, the applicability of new oxygen-sensitive nanoparticles operated within the more suitable infrared wavelength region are introduced and validated for DOT measurement.ResultsThe biocompatibility of the used dispersed oxygen-sensitive nanoparticles was proven via RAMOS cultivations for Hansenula polymorpha, Gluconobacter oxydans, and Escherichia coli. The applicability of the introduced DOT measurement technique for online monitoring of cultivations was demonstrated and successfully validated. The nanoparticles showed no disturbing effect on the online measurement of the fluorescence intensities of the proteins GFP, mCherry and YFP measured by a BioLector prototype. Additionally, the DOT measurement was not influenced by changing concentrations of these proteins. The kLa values for the applied cultivation conditions were successfully determined based on the measured DOT.ConclusionsThe introduced technique appeared to be practically as well as economically advantageous for DOT online measuring in microtiter plates. The disadvantage of limited availability of microtiter plates with immobilized sensor spots (optodes) does not apply for this introduced technique. Due to the infrared wavelength range, used for the DOT measurement, no interferences with biogenic fluorescence or with expressed fluorescent proteins (e.g. YFP, GFP or mCherry) occur.

【 授权许可】

CC BY   
© Ladner et al. 2015

【 预 览 】
附件列表
Files Size Format View
RO202311103369912ZK.pdf 2131KB PDF download
MediaObjects/12951_2023_2137_MOESM1_ESM.docx 441KB Other download
Fig. 1 1367KB Image download
MediaObjects/12888_2023_5253_MOESM1_ESM.docx 105KB Other download
Fig. 4 1643KB Image download
12951_2016_225_Article_IEq3.gif 1KB Image download
Fig. 6 889KB Image download
12951_2016_225_Article_IEq2.gif 1KB Image download
Fig. 2 159KB Image download
【 图 表 】

Fig. 2

12951_2016_225_Article_IEq2.gif

Fig. 6

12951_2016_225_Article_IEq3.gif

Fig. 4

Fig. 1

【 参考文献 】
  • [1]
  • [2]
  • [3]
  • [4]
  • [5]
  • [6]
  • [7]
  • [8]
  • [9]
  • [10]
  • [11]
  • [12]
  • [13]
  • [14]
  • [15]
  • [16]
  • [17]
  • [18]
  • [19]
  • [20]
  • [21]
  • [22]
  • [23]
  • [24]
  • [25]
  • [26]
  • [27]
  • [28]
  • [29]
  • [30]
  • [31]
  • [32]
  • [33]
  • [34]
  • [35]
  • [36]
  • [37]
  文献评价指标  
  下载次数:11次 浏览次数:2次