| Improved accuracy and precision of fat-suppressed isotropic 3D T2 mapping MRI of the knee with dictionary fitting and patch-based denoising | |
| Original Article | |
| Roberto Colotti1  Julien Favre2  Aicha Lamri-Senouci3  Simone Rumac3  Simon Kuhn3  Ruud B. van Heeswijk3  Patrick Omoumi3  Jean-Baptiste Ledoux4  Jérôme Yerly4  Aurélien Bustin5  Jessica A. M. Bastiaansen6  | |
| [1] Biomedical Data Science Center (BDSC), Lausanne University Hospital (CHUV) and University of Lausanne (UNIL), Lausanne, Switzerland;Department of Musculoskeletal Medicine, Lausanne University Hospital (CHUV) and University of Lausanne (UNIL), Lausanne, Switzerland;The Sense Innovation and Research Center, Lausanne and Sion, Switzerland;Department of Radiology, Lausanne University Hospital (CHUV) and University of Lausanne (UNIL), Lausanne, Switzerland;Department of Radiology, Lausanne University Hospital (CHUV) and University of Lausanne (UNIL), Lausanne, Switzerland;Center for BioMedical Imaging (CIBM), Lausanne, Switzerland;Department of Radiology, Lausanne University Hospital (CHUV) and University of Lausanne (UNIL), Lausanne, Switzerland;Department of Cardiovascular Imaging, Hôpital Cardiologique du Haut-Lévêque, CHU de Bordeaux, France;IHU LIRYC, Electrophysiology and Heart Modeling Institute, INSERM U1045, Centre de Recherche Cardio-Thoracique de Bordeaux, Université de Bordeaux, Bordeaux, France;Department of Radiology, Lausanne University Hospital (CHUV) and University of Lausanne (UNIL), Lausanne, Switzerland;Department of Diagnostic, Interventional and Pediatric Radiology (DIPR), Bern University Hospital, University of Bern, Inselspital, Switzerland;Translation Imaging Center (TIC), Swiss Institute for Translational and Entrepreneurial Medicine, Bern, Switzerland; | |
| 关键词: Cartilage; Knee joint; Magnetic resonance imaging; Phantoms (imaging); | |
| DOI : 10.1186/s41747-023-00339-8 | |
| received in 2022-12-21, accepted in 2023-03-15, 发布年份 2023 | |
| 来源: Springer | |
PDF
|
|
【 摘 要 】
PurposeTo develop an isotropic three-dimensional (3D) T2 mapping technique for the quantitative assessment of the composition of knee cartilage with high accuracy and precision.MethodsA T2-prepared water-selective isotropic 3D gradient-echo pulse sequence was used to generate four images at 3 T. These were used for three T2 map reconstructions: standard images with an analytical T2 fit (AnT2Fit); standard images with a dictionary-based T2 fit (DictT2Fit); and patch-based-denoised images with a dictionary-based T2 fit (DenDictT2Fit). The accuracy of the three techniques was first optimized in a phantom study against spin-echo imaging, after which knee cartilage T2 values and coefficients of variation (CoV) were assessed in ten subjects in order to establish accuracy and precision in vivo. Data given as mean ± standard deviation.ResultsAfter optimization in the phantom, whole-knee cartilage T2 values of the healthy volunteers were 26.6 ± 1.6 ms (AnT2Fit), 42.8 ± 1.8 ms (DictT2Fit, p < 0.001 versus AnT2Fit), and 40.4 ± 1.7 ms (DenDictT2Fit, p = 0.009 versus DictT2Fit). The whole-knee T2 CoV reduced from 51.5% ± 5.6% to 30.5 ± 2.4 and finally to 13.1 ± 1.3%, respectively (p < 0.001 between all). The DictT2Fit improved the data reconstruction time: 48.7 ± 11.3 min (AnT2Fit) versus 7.3 ± 0.7 min (DictT2Fit, p < 0.001). Very small focal lesions were observed in maps generated with DenDictT2Fit.ConclusionsImproved accuracy and precision for isotropic 3D T2 mapping of knee cartilage were demonstrated by using patch-based image denoising and dictionary-based reconstruction.Key points• Dictionary T2 fitting improves the accuracy of three-dimensional (3D) knee T2 mapping.• Patch-based denoising results in high precision in 3D knee T2 mapping.• Isotropic 3D knee T2 mapping enables the visualization of small anatomical details.
【 授权许可】
CC BY
© The Author(s) 2023
【 预 览 】
| Files | Size | Format | View |
|---|---|---|---|
| RO202308153093518ZK.pdf | 2230KB | ||
| Fig. 1 | 424KB | Image | |
| 41116_2023_36_Article_IEq153.gif | 1KB | Image | |
| 41116_2023_36_Article_IEq154.gif | 1KB | Image | |
| Fig. 1 | 252KB | Image | |
| 41116_2023_36_Article_IEq191.gif | 1KB | Image | |
| 41116_2023_36_Article_IEq202.gif | 1KB | Image | |
| Fig. 3 | 183KB | Image | |
| Fig. 1 | 129KB | Image | |
| 41116_2023_36_Article_IEq209.gif | 1KB | Image | |
| 41116_2023_36_Article_IEq210.gif | 1KB | Image | |
| 41116_2023_36_Article_IEq211.gif | 1KB | Image |
【 图 表 】
41116_2023_36_Article_IEq211.gif
41116_2023_36_Article_IEq210.gif
41116_2023_36_Article_IEq209.gif
Fig. 1
Fig. 3
41116_2023_36_Article_IEq202.gif
41116_2023_36_Article_IEq191.gif
Fig. 1
41116_2023_36_Article_IEq154.gif
41116_2023_36_Article_IEq153.gif
Fig. 1
【 参考文献 】
- [1]
- [2]
- [3]
- [4]
- [5]
- [6]
- [7]
- [8]
- [9]
- [10]
- [11]
- [12]
- [13]
- [14]
- [15]
- [16]
- [17]
- [18]
- [19]
- [20]
- [21]
- [22]
- [23]
- [24]
- [25]
- [26]
- [27]
PDF