Evaluation of beam hardening artifacts around dental implants: CT study on bovine ribs | |
article | |
Serindere Gozde1  Aktuna Belgin Ceren1  Serindere Mehmet2  Berkay Belgin Huseyin3  Orhan Kaan4  | |
[1] Hatay Mustafa Kemal University, Faculty of Dentistry, Department of Dentomaxillofacial Radiology;Hatay Education and Research Hospital, Department of Radiology;Private Dental Clinic;Ankara University, Faculty of Dentistry, Department of Dentomaxillofacial Radiology, Ankara, Turkey + Ankara University, Medical Design Application and Research Center | |
关键词: beam hardening artifact; bone density; computed tomography; dental implant; | |
DOI : 10.5937/bjdm2301028S | |
学科分类:社会科学、人文和艺术(综合) | |
来源: Walter de Gruyter GmbH | |
【 摘 要 】
Background/Aim: The aim of this study was to evaluate beam hardening artifacts generated by Grade 4 and Grade 5 dental implants on computed tomography (CT) images at low and high kilovoltage peaks (kVp). Material and Methods: A total of 16 implants, 8 of which were Grade 4 and 8 were Grade 5, were inserted into bovine ribs. CT images of bovine ribs were acquired using two different exposure protocol: low kVp and high kVp. Beam hardening artifacts generated by Grade 4 and Grade 5 dental implants were calculated by the mean Hounsfield unit (HU) within a standardized region-of-interest (ROI). Results: Artifact in Grade 4 implants were greater than that in Grade 5 implants. Also, artifacts at the high kVp were lower than that at the low kVp. Conclusions: CT scans providing HU values can be used to evaluate the beam hardening artifact. Beam hardening artifacts decreased in the CT images with high kVp. Grade 5 dental implants have an advantage by producing less severe beam hardening artifacts.
【 授权许可】
CC BY
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
RO202307090004263ZK.pdf | 501KB | download |