EJNMMI Physics | |
Investigation of image-based lesion and kidney dosimetry protocols for 177Lu-PSMA-I&T therapy with and without a late SPECT/CT acquisition | |
Original Research | |
Sandra Resch1  Astrid Delker1  Mikhail Rumiantcev1  Lena M. Unterrainer1  Mathias Zacherl1  Gabriel T. Sheikh1  Peter Bartenstein1  Grigory Liubchenko1  Sarah Takayama Fouladgar1  Harun Ilhan1  Vera Wenter1  Guido Böning1  Leonie Beyer1  Sibylle I. Ziegler1  | |
[1] Department of Nuclear Medicine, University Hospital, LMU Munich, Munich, Germany; | |
关键词: SPECT; PSMA; Lu; Prostate cancer; Dosimetry; | |
DOI : 10.1186/s40658-023-00529-8 | |
received in 2022-09-27, accepted in 2023-01-26, 发布年份 2023 | |
来源: Springer | |
【 摘 要 】
Background177Lu-PSMA therapy has been successfully used to prolong the survival of patients with metastatic castration-resistant prostate cancer. Patient-specific dosimetry based on serial quantitative SPECT/CT imaging can support the understanding of dose–effect relationships. However, multiple SPECT/CT measurements can be challenging for patients, which motivates the investigation of efficient sampling schedules and their impact on dosimetry. In this study, different time samplings with respect to the number and timing of SPECT/CT acquisitions with and without a late measurement were investigated.Materials and methodsIn total, 43 lesions and 10 kidneys of 5 patients receiving 177Lu-PSMA-I&T therapy were investigated. Whole-body SPECT/CT measurements were performed at 1, 2, 3 and 7 days post-injection. For both lesions (isocontour-based segmentation) and kidneys (CT-based segmentation), a reference model was employed including all four time points. To identify the best-matching fit function out of a pre-defined set of models, visual inspection, coefficients of variation and sum of squared errors were considered as goodness-of-fit criteria. Biologically effective doses (BEDs) calculated with different time samplings (days 1, 2, 3/1, 2, 7/1, 3, 7/2, 3, 7 and 1, 2/1, 3/1, 7) were compared to the reference.ResultsThe best-fit function was found to be a mono-exponential model for lesions and a bi-exponential model with a population-based parameter and two free parameters for kidneys. The BEDs calculated with the time sampling 1, 3, 7 days showed the lowest deviations from the reference for lesions with 4 ± 5%. Without day 7, still 86% of all lesions showed deviations from the reference < 10%. The outlier deviations showed a positive correlation with the effective half-life of the respective lesions. For kidneys, including days 1, 2, 3 achieved the best results with 0 ± 1%. Generally, deviations for kidneys were found to be small for all time samplings (max. 13%).ConclusionsFor combined optimization of the SPECT/CT time sampling for kidney and lesion dosimetry during 177Lu-PSMA-I&T therapy, the sampling with days 1, 3, 7 showed the smallest deviation from the reference. Without a late acquisition, using the schedule with days 1, 2, 3 is likewise feasible.
【 授权许可】
CC BY
© The Author(s) 2023
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
RO202305159414169ZK.pdf | 1233KB | download | |
40517_2023_248_Article_IEq14.gif | 1KB | Image | download |
MediaObjects/41408_2023_791_MOESM6_ESM.xlsx | 11KB | Other | download |
MediaObjects/12974_2023_2735_MOESM1_ESM.docx | 2084KB | Other | download |
Fig. 2 | 1316KB | Image | download |
Fig. 1 | 1688KB | Image | download |
【 图 表 】
Fig. 1
Fig. 2
40517_2023_248_Article_IEq14.gif
【 参考文献 】
- [1]
- [2]
- [3]
- [4]
- [5]
- [6]
- [7]
- [8]
- [9]
- [10]
- [11]
- [12]
- [13]
- [14]
- [15]
- [16]
- [17]
- [18]
- [19]
- [20]
- [21]
- [22]
- [23]
- [24]
- [25]
- [26]
- [27]
- [28]
- [29]
- [30]
- [31]
- [32]
- [33]
- [34]
- [35]
- [36]
- [37]
- [38]
- [39]
- [40]
- [41]