Accuracy of atrial fibrillation detection by an insertable cardiac monitor in patients undergoing catheter ablation: Results of the BioVAD study | |
article | |
Amira Assaf MD1  Dominic A. M. J. Theuns PhD1  Rafi Sakhi MD, PhD1  Rohit E. Bhagwandien MD1  Tamas Szili-Torok MD, PhD1  Sing-Chien Yap MD, PhD1  | |
[1] Department of Cardiology, Erasmus MC, University Medical Center Rotterdam | |
关键词: atrial fibrillation; BioMonitor III; catheter ablation; implantable loop recorder; insertable cardiac monitor; pulmonary vein isolation; | |
DOI : 10.1111/anec.12960 | |
来源: Wiley | |
【 摘 要 】
Background Insertable cardiac monitors (ICMs) are increasingly used to evaluate the atrial fibrillation (AF) burden after catheter ablation of AF. BioMonitor III (BM3) is an ICM with a long sensing vector, which enhances sensing capabilities. The AF detection algorithm of the BM3 is based on R–R interval variability. Objective To evaluate the performance of the AF detection algorithm of BM3 in patients before and after catheter ablation of AF using simultaneous Holter recordings. Methods In this prospective study, we enrolled patients scheduled for catheter ablation of paroxysmal or persistent AF. After BM3 implantation, patients had a 4 days Holter registration before and 3 months after ablation. All true AF episodes ≥2 min on the Holter were annotated and matched with BM3 detected AF detections. Results Thirty-one patients were enrolled (mean age 60 ± 8, 74% male, 68% paroxysmal AF). Fifty-six Holter registrations were performed in 30 patients. Twelve patients demonstrated at least one true AF episode with a total AF duration of 570 h. The AF burden accuracy of BM3 before catheter ablation was 99.6%, with a duration sensitivity of 98.6% and a duration specificity of 99.9%. The AF burden accuracy of BM3 after catheter ablation was 99.8%, with a duration sensitivity of 90.2% and a duration specificity of 99.9%. Overall, the AF burden detected on the Holter and BM3 demonstrated a high Pearson correlation coefficient of 0.996. Conclusion BM3 accurately detects AF burden in patients before and after catheter ablation of AF.
【 授权许可】
Unknown
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
RO202302050004560ZK.pdf | 1476KB | download |