期刊论文详细信息
JACC: Case Reports
Electrostatic Discharge Causing Pump Shutdown in HeartMate 3
Tonje Thorvaldsen, MD, PhD1  Ann Hallberg Kristensen2  Lars H. Lund, MD, PhD2  Emil Najjar, MD, PhD3  Ulrich P. Jorde, MD, PhD3  Magnus Dalén, MD, PhD4 
[1] Address for correspondence: Dr. Emil Najjar, Karolinska Institutet, Heart and Vascular Theme Karolinska University Hospital, N305, 17176 Stockholm, Sweden.;Heart and Vascular Theme, Karolinska University Hospital, Stockholm, Sweden;Department of Medicine, Karolinska Institutet, Stockholm, Sweden;Heart and Vascular Theme, Karolinska University Hospital, Stockholm, Sweden;
关键词: complications;    device;    device malfunction;    heart failure;    HeartMate 3;    left ventricular assist device;   
DOI  :  
来源: DOAJ
【 摘 要 】

Left ventricular assist devices (LVADs) improve symptoms and outcomes in advanced heart failure. Although device malfunction has decreased significantly with later generation LVADs, it has not been eliminated. We describe the clinical course of a patient with HeartMate 3 LVAD who experienced device malfunction, involving temporary pump shutdown suspected to be caused by electrostatic discharge. (Level of Difficulty: Advanced.)

【 授权许可】

Unknown   

  文献评价指标  