Development of a Photoelectric Adjustment System With Extended Range for Fluorescence Immunochromatographic Assay Strip Readers | |
Min Du1  Yueming Gao2  Huihuang Wu2  Siohang Pun2  Jiejie Yang2  Mangi Vai3  | |
[1] Pharmaceutical Technology of Fujian Province, Fuzhou, China;College of Physics and Information Engineering, Fuzhou University, Fuzhou, China;;Key Lab of Medical Instrumentation & | |
关键词: Fluorescence immunochromatographic assay; photoelectric adjustment; light intensity; detection range; | |
DOI : 10.1109/JPHOT.2021.3075900 | |
来源: DOAJ |
【 摘 要 】
Fluorescence immunochromatographic assay (FICA) is a quantitative detection technique widely used in clinical diagnosis, environmental monitoring, and food safety. To deal with the limited applications caused by insufficient detection range, this paper proposes a photoelectric adjustment system suitable for FICA strip readers to expand its detection range. The photoelectric adjustment system is proposed based on the relationship between the excitation light intensity and the fluorescence intensity, which provides the optimal excitation light intensity and a stable baseline amplitude for the FICA strip reader. To verify the proposed system, it was applied to the strip reader we had previously developed. The results show that the linear detection range of the FICA strip reader was extended from 1.95-256 μg/mL to 1-1024 μg/mL after using the proposed method. At the same time, the accuracy of the FICA strip reader does not deteriorate and is in good comparisons with the conventional ESEQuant Lateral Flow Reader (with