| Sensors | |
| An On-Orbit Dynamic Calibration Method for an MHD Micro-Angular Vibration Sensor Using a Laser Interferometer | |
| Fan Liu1  Ganmin Xia1  Yingjie Wu1  Xingfei Li1  | |
| [1] State Key Laboratory of Precision Measuring Technology and Instruments, Tianjin University, Tianjin 300072, China; | |
| 关键词: mhd; micro-angular vibration sensor; on-orbit dynamic calibration; sensitivity; laser interferometer; installation error; | |
| DOI : 10.3390/s19194291 | |
| 来源: DOAJ | |
【 摘 要 】
The magnetohydrodynamic (MHD) micro-angular vibration sensor is a significant component of the MHD Inertial Reference Unit (MIRU) and measures micro-amplitude and wide frequency angular vibration. The MHD micro-angular vibration sensor must be calibrated in orbit since the ground calibration parameters may change after lift-off. An on-orbit dynamic calibration method for the MHD micro-angular vibration sensor is proposed to calibrate the complex sensitivity of the sensor in high frequency. An absolute calibration method that combines a homodyne laser interferometer and an angular retroreflector was developed. The sinusoidal approximation method was applied, and the calibration system was established and tested using a manufactured MHD sensor. Furthermore, the measurement principle and installation errors were analyzed, including the eccentric installation error of the retroreflector, the tilt installation error of the retroreflector, and the optical path tilt error. This method can be realized within a rotation range of
【 授权许可】
Unknown