Insights into Imaging | |
Dynamic MRI for articulating joint evaluation on 1.5 T and 3.0 T scanners: setup, protocols, and real-time sequences | |
Marc Garetier1  Bhushan Borotikar2  Karim Makki2  François Rousseau2  Douraïed Ben Salem2  Sylvain Brochard2  | |
[1] Department of Radiology, Military Teaching Hospital Clermont-Tonnerre;Laboratory of Medical Information Processing (LATIM), INSERM-UMR 1101; | |
关键词: Joints; Motion; Musculoskeletal system; Magnetic resonance imaging; | |
DOI : 10.1186/s13244-020-00868-5 | |
来源: DOAJ |
【 摘 要 】
Abstract Dynamic magnetic resonance imaging (MRI) is a non-invasive method that can be used to increase the understanding of the pathomechanics of joints. Various types of real-time gradient echo sequences used for dynamic MRI acquisition of joints include balanced steady-state free precession sequence, radiofrequency-spoiled sequence, and ultra-fast gradient echo sequence. Due to their short repetition time and echo time, these sequences provide high temporal resolution, a good signal-to-noise ratio and spatial resolution, and soft tissue contrast. The prerequisites of the evaluation of joints with dynamic MRI include suitable patient installation and optimal positioning of the joint in the coil to allow joint movement, sometimes with dedicated coil support. There are currently few recommendations in the literature regarding appropriate protocol, sequence standardizations, and diagnostic criteria for the use of real-time dynamic MRI to evaluate joints. This article summarizes the technical parameters of these sequences from various manufacturers on 1.5 T and 3.0 T MRI scanners. We have reviewed pertinent details of the patient and coil positioning for dynamic MRI of various joints. The indications and limitations of dynamic MRI of joints are discussed.
【 授权许可】
Unknown