The stability of radiance measurements taken by the Sky Quality Meter (SQM)was tested under rapidly changing temperature conditions during exposure to a stable lightfield in the laboratory. The reported radiance was found to be negatively correlated withtemperature, but remained within 7% of the initial reported radiance over a temperaturerange of -15 °C to 35 °C, and during temperature changes of -33 °C/h and +70 °C/h.This is smaller than the manufacturer’s quoted unit-to-unit systematic uncertainty of 10%,indicating that the temperature compensation of the SQM is adequate under expected outdoor operating conditions.