Temperature Stability of the Sky Quality Meter
Sabrina Schnitt1  Thomas Ruhtz1  Jürgen Fischer1  Franz Hölker2 
[1] Institute for Space Sciences, Freie Universität Berlin, Carl-Heinrich-Becker-Weg 6-10, Berlin 12165, Germany; E-Mails:;Leibniz-Institute of Freshwater Ecology and Inland Fisheries, Müggelseedamm 310, Berlin 12587, Germany; E-Mail:
关键词: light pollution;    sky quality meter;    light at night;    temperature;    stability;    skyglow;    night sky brightness;   
DOI  :  10.3390/s130912166
来源: mdpi
PDF
【 摘 要 】

The stability of radiance measurements taken by the Sky Quality Meter (SQM) was tested under rapidly changing temperature conditions during exposure to a stable light field in the laboratory. The reported radiance was found to be negatively correlated with temperature, but remained within 7% of the initial reported radiance over a temperature range of −15°C to 35°C, and during temperature changes of −33°C/h and +70°C/h. This is smaller than the manufacturer's quoted unit-to-unit systematic uncertainty of 10%, indicating that the temperature compensation of the SQM is adequate under expected outdoor operating conditions.

【 授权许可】

CC BY   
© 2013 by the authors; licensee MDPI, Basel, Switzerland.

【 预 览 】
附件列表
Files Size Format View
RO202003190033237ZK.pdf 3570KB PDF download
  文献评价指标  