The Effects of Silicone Enclosure Colour on the Function of Optical Sensors | |
Shahbaz Askari1  Sadra Khosravi1  Babak Shadgan1  Garrett Frank1  Katharina Raschdorf1  Brian K. Kwon1  | |
[1] International Collaboration on Repair Discoveries, 818 West 10th Avenue, Vancouver, BC V5Z 1M9, Canada; | |
关键词: implantable; biosensor; phantom; diagnostics; direct measurement; absorption; | |
DOI : 10.3390/biology11060932 | |
来源: DOAJ |
【 摘 要 】
The colour of the silicone enclosure of an implantable reflectance-based optical probe plays a critical role in sensor performance. Red-coloured probes that are highly reflective to near-infrared light have been found to increase photodetector power by a factor of 6 for wavelengths between 660 and 950 nm and triple the magnitude of measured cardiac pulsations compared to traditional black probes. The increase in photodetector power and cardiac pulsation magnitude is presumably due to increased spatial range resulting from a higher magnitude of superficial tissue scattering. Conversely, probes with highly absorbent colours such as black and blue result in more stable signals and are expected to have higher spatial resolution and depth of penetration.