期刊论文详细信息
The Journal of Engineering
Estimation of pulsatile flow and differential pressure based on multi-layer perceptron using an axial flow blood pump
Tan Jianping1  Wu Weiqiang1  Dang Caixin1  Wu Kun1  Yu Zheqin1  Wang Shuai1 
[1] School of Mechanical and Electrical Engineering, Central South University;
关键词: blood vessels;    blood pressure measurement;    pulsatile flow;    haemodynamics;    multilayer perceptrons;    blood;    pumps;    pulsatile flow;    differential pressure;    multilayer perceptron;    axial flow blood pump;    noninvasive method;    pulsating flow;    estimation model;    pulsating conditions;    pulsation waveform;    flow error;    pressure difference error;    traditional flow;    pressure prediction method;   
DOI  :  10.1049/joe.2020.0045
来源: DOAJ
【 摘 要 】

This study proposes a non-invasive method for estimating the pulsating flow and pressure difference, which uses the blood pump estimation model based on a multi-layer perceptron to calculate the flow and pressure difference under pulsating conditions. The model takes 11 parameters such as the rotational speed, power, and pulsation waveform of the blood pump as the input and uses the pressure difference and flow as the output. The experimental results of 119,590 sample data show that the flow error of the training set of the blood pump estimation model is 0.14 l/min and the pressure difference error is 7.50 mmHg; the flow error of the test set is 0.14 l/min and the pressure difference error is 7.50 mmHg. Compared with the traditional flow and pressure prediction method, this method has higher precision, which will provide a certain technical accumulation for accurately estimating the flow and pressure difference of the blood pump in the pulsating conditions.

【 授权许可】

Unknown   

  文献评价指标  
  下载次数:0次 浏览次数:0次