Nuclear Technology and Radiation Protection
Patient doses and image quality in chest radiography: The influence of different beam qualities
关键词: chest radiography;    optimization;    dose;    image quality;   
DOI  :  10.2298/NTRP0702048C
来源: DOAJ
【 摘 要 】

A simple method of assessing optimal X-ray beam quality in respect to patient exposure and image quality in chest screen-film radiography is presented here. Different beam qualities were generated by the use of various combinations of tube voltages (70 kV to 110 kV) and Al and Cu filter thick nesses. Patient doses were assessed by kerma-area product measurements. Simultaneously, image quality was evaluated by a twofold method: a clinical study applying European quality criteria for the radiographic technique of image on image of 126 patients and a multifunctional home-made dosimetric phantom with embedded test objects. The quantification of image quality criteria yields a simpler method of optimizing image quality and patient dose relationships. Modifications of radiographic practice, based on image quality assessment and dose measurements, resulted in significant dose reductions and preservation of image quality. Through the use of harder beam quality, dose reduction of up to a value of factor 3 were observed, compared to the doses from previously used radiographic techniques, implying that sufficient image quality does not necessarily imply higher doses. As a result of the optimization process, an optimal radiographic technique was suggested.