期刊论文详细信息
Applied Sciences
The Application of Adaptive Time Gain Compensation in an Improved Breast Ultrasound Tomography Algorithm
Yijun Cheng1  Guojun Zhang1  Binzhen Zhang1  Chang Liu2  Chenyang Xue2  Wendong Zhang2 
[1] Dynamic Measurement, North University of China, Ministry of Education, Taiyuan 030051, China;;Key Laboratory of Instrumentation Science &
关键词: ultrasound tomography;    attenuation coefficient;    TGC;    PMUT linear array;   
DOI  :  10.3390/app9122574
来源: DOAJ
【 摘 要 】

In order to better detect information about a mass in breast tissue, an ultrasound tomography algorithm based on adaptive time gain compensation (TGC) was designed. Field II was utilized to automatically evaluate the phantom attenuation coefficient and compensate for the attenuated image. The image reconstruction algorithm process is presented here. Furthermore, the experimental setup with the cylindrical motion of a piezoelectric micromachined ultrasonic transducer (PMUT) linear array was used to detect the mass in the breast model. The attenuation coefficient was evaluated by using the spectral cross-correlation method. According to the acquired attenuation coefficients, TGC compensates for the pulse-echo signal, and the horizontal slice image was reconstructed using the tomography algorithm. The experimental results show that this algorithm can evaluate the attenuation coefficient of the breast model and improve the ability to detect an internal mass. At the same time, the realization of attenuation compensation with software is beneficial to the development of portable medical equipment.

【 授权许可】

Unknown   

  文献评价指标  
  下载次数:0次 浏览次数:0次