| Model eyes with curved multilayer structure for the axial resolution evaluation of an ophthalmic optical coherence tomography device | |
| Zhenggang Cao1  Zhixiong Hu1  Wenli Liu1  Zengqian Ding2  Wen Qiao2  Xiaojun Chen3  | |
| [1] National Institute of Metrology, Division of Medical and Biological Measurements, Beijing 100029, P. R. China;Optoelectronics and Energy & Collaborative Innovation, Center of Suzhou Nano Science and Technology College of Physics, Soochow University, Suzhou 215006, P. R. China;School of Mechanical Engineering, Shanghai Jiao Tong University, Shanghai 200240, P. R. China; | |
| 关键词: Optical coherence tomography; metrology; optical standards and testing; spin coating; | |
| DOI : 10.1142/S179354581850013X | |
| 来源: DOAJ | |
【 摘 要 】
Optical coherence tomography (OCT) has been widely applied to the diagnosis of eye diseases during the past two decades. However, valid evaluation methods are still not available for the clinical OCT devices. In order to assess the axial resolution of the OCT system, standard model eyes with micro-scale multilayer structure have been designed and manufactured in this study. Mimicking a natural human eye, proper Titanium dioxide (TiO2) materials of particles with different concentrations were selected by testing the scattering coefficient of PDMS phantoms. The artificial retinas with multilayer films were fabricated with the thicknesses from 9.5 to 30 micrometers using spin coating technology. Subsequently, standard OCT model eyes were accomplished by embedding the retina phantoms into the artificial frames of eyes. For ease of measurement processing, a series of model eyes were prepared, and each contained films with three kinds of thicknesses. Considering the traceability and accuracy of the key parameters of the standard model eyes, the thicknesses of multilayer structures were verified using Thickness Monitoring System. Through the experiment with three different OCT devices, it demonstrated the model eyes fabricated in this study can provide an effective evaluation method for the axial resolution of an ophthalmic OCT device.
【 授权许可】
Unknown