The usefulness of three-dimensional ultrasound fusion imaging for precise needle placement in liver thermal ablation: a phantom and an in vivo simulation study
Ming Kuang1  Yanling Zheng2  Xiaoyan Xie2  Manying Li2  Xiaohua Xie2  Xiaoer Zhang2  Ming Xu2  Guangliang Huang2  Yuqing Guo2  Kaixin Yu3 
[1] Department of Liver Surgery, Division of Interventional Ultrasound, the First Affiliated Hospital of Sun Yat-sen University, Guangzhou, China;Division of Interventional Ultrasound, Department of Medical Ultrasonics, Institute of Diagnostic and Interventional Ultrasound, the First Affiliated Hospital of Sun Yat-sen University, Guangzhou, China;Medical Imaging System Division, Shenzhen Mindray Bio-Medical Electronics Co. Ltd, Shenzhen, China;
关键词: Three-dimensional ultrasound;    fusion imaging;    needle placement;    thermal ablation;    precise;    liver tumors;   
DOI  :  10.1080/02656736.2022.2057596
来源: DOAJ
【 摘 要 】

Purpose To investigate the value of three-dimensional ultrasound fusion imaging (3DUS-FI) in real-time guiding needle placement by phantom models and in vivo simulations.Materials and methods Two radiologists (beginner and expert) performed needle placement using two-dimensional ultrasound (2DUS) and 3DUS-FI, respectively. In the phantom study, single-needle placement was performed by puncturing the center point of each ball and assessed based on the specimen length. Multiple-needles placement was performed by placing three needles in each ball, and their locations were confirmed by computed tomography, and assessed based on the distance deviation between needles. In the in vivo simulation study, simulated-needle placement was performed by placing a virtual ablation needle in each liver tumor and assessed by the simulated ablative cover rate and margin.Results Specimen length was significantly longer with 3DUS-FI in the beginner, whereas no significant difference was observed in the expert (2DUS vs. 3DUS-FI: beginner, 14.60 ± 2.60 mm vs. 16.25 ± 1.38 mm, p = .017; expert, 16.78 ± 1.40 mm vs. 16.95 ± 1.15 mm, p = .668). Distance deviation between needles was significantly smaller with 3DUS-FI (2DUS vs. 3DUS-FI: beginner, 25.06 ± 16.07 mm vs. 3.72 ± 1.99 mm, p < .001; expert, 11.70 ± 7.79 mm vs. 2.89 ± 1.52 mm, p < .001). The simulated ablative cover rate and margin were significantly larger with 3DUS-FI for the beginner, whereas only the latter was significantly larger for the expert (2DUS vs. 3DUS-FI: beginner, 73.55 ± 8.73% vs. 81.38 ± 11.84%, p = .001, 0.82 ± 0.97 mm vs. 2.65 ± 1.23 mm, p < .001; expert, 78.60 ± 9.91% vs. 83.24 ± 11.69%, p = .059; 1.65 ± 1.15 mm vs. 2.95 ± 1.13 mm, p < .001).Conclusions 3DUS-FI is useful for real-time guiding precise needle placement and may be further use to improve the efficacy of liver thermal ablation.

【 授权许可】

Unknown   

  文献评价指标  
  下载次数:0次 浏览次数:0次