期刊论文详细信息
Current Directions in Biomedical Engineering
tISM - a transportable integrating sphere setup for standardized measurements of blood and tissue phantoms
Gehring Hartmut1  Wegerich Philipp2  Rackebrandt Klaas3 
[1] Departement of Anesthesiology and Intensive Care Medicine, University Medical Center, Ratzeburger Allee 160 23562 Lübeck, Germany;Institute of Medical Engineering and Graduate School for Computing in Medicine and Life Sciences, Universität zu Lübeck, Ratzeburger Allee 160 23562 Lübeck, Germany;Institute of Medical Engineering, Universität zu Lübeck, Ratzeburger Allee 160 23562 Lübeck, Germany;
关键词: phantom model;    optical-interaction-coefficients;    intralipid;    blood;   
DOI  :  10.1515/cdbme-2017-0061
来源: DOAJ
【 摘 要 】

The development of an optical sensor for non-invasive measurements in humans requires a test setup, where the optical properties of tissue and blood can be adjusted and measured standardized. The goal of this work is to develop a simplified device based on an integrating sphere setup to evaluate the optical properties of tissue and blood phantoms with respect to static as well as flow conditions. Furthermore, the measurement system is intended to be used at different locations such as laboratories and operating theaters. We evaluate the absorption μa and reduced scattering μs' coefficients of specimens, with the developed integrating sphere setup. The measurement is regulated by a microcontroller for averaging and processing the data. The system is housed in a lightproof box and powered by a battery and therefore transportable. Due to this construction, no calibration is necessary between transports of the system. Calculations are executed with the inverse adding doubling algorithm. In order to basically calibrate and evaluate the setup before first transportation, a dilution series with Intralipid and India ink serve for the test. The results were consistent with precedent studies (mean absolute deviation for μs' of 0.75 mm-1) and demonstrate that this method might be able to produce liquids with adjustable optical properties, as required for further research. Furthermore, a first dilution series of heparinized heamoglobin (5 to 15 g/dl) with oxygen saturation of 98 % was measured with this system under flow conditions.

【 授权许可】

Unknown   

  文献评价指标  
  下载次数:0次 浏览次数:3次