Using additional pressure control lines when connecting a continuous renal replacement therapy device to an extracorporeal membrane oxygenation circuit
Chi Ryang Chung1  Soo Jin Na1  Kyeongman Jeon1  Gee Young Suh1  Yang Hyun Cho2  Hye Ryoun Jang3  Hee Jung Choi4 
[1] Department of Critical Care Medicine, Samsung Medical Center, Sungkyunkwan University School of Medicine;Department of Thoracic and Cardiovascular Surgery, Samsung Medical Center, Sungkyunkwan University School of Medicine;Division of Nephrology, Department of Medicine, Samsung Medical Center, Sungkyunkwan University School of Medicine;Intensive Care Unit Nursing Department, Samsung Medical Center, Sungkyunkwan University School of Medicine;
关键词: Renal replacement therapy;    Extracorporeal membrane oxygenation;    Critical care;    Acute kidney injury;   
DOI  :  10.1186/s12882-018-1172-2
来源: DOAJ
【 摘 要 】

Abstract Background The introduction of a continuous renal replacement therapy (CRRT) device into the extracorporeal membrane oxygenation (ECMO) circuit is widely used. However, excessive pressure transmitted to the CRRT device is a major disadvantage. We investigated the effects of using additional pressure control lines on the pressure and the lifespan of the CRRT circuit connected to the ECMO. Methods This is an observational study using prospectively collected data from consecutive patients receiving CRRT connected into the ECMO circuit at a university-affiliated, tertiary hospital from January 2013 to December 2016. The CRRT circuit was connected into the ECMO circuit through the Luer Lock connection without an additional pressure control line in 16 patients (9%, no line group), an additional pressure control line on the inlet line in 36 patients (23%, single line group), and additional pressure control lines on both the inlet and outlet lines in 118 patients (77%, double line group). The outcome measures of interest were compared among the three groups. Results The median access pressure was higher in the no line group compared to the groups. However, median filter pressure, effluent pressure, and return pressure were higher in the double line group compared to the other groups. There were no significant differences in platelets, lactate dehydrogenase, and plasma hemoglobin among the 3 groups over the time period studied. Median lifespan of the CRRT circuits in the double line group was 45.0 (29.0–63.7) hours, which was higher compared to 21.8 (11.6–31.8) hours in the no line group and 23.0 (15.0–34.6) hours in the single line group, respectively. In addition, in-hospital mortality was lower in the double line group (48.3%) compared to the no line group (68.8%) and the single line group (75.0%). Conclusions Additional tubing can be considered a simple and safe method for pressure control and lengthening circuit survival when connecting the CRRT device to the ECMO circuit.

【 授权许可】

Unknown   

  文献评价指标  
  下载次数:0次 浏览次数:9次