Design, characterization and evaluation of a laser-guided focused ultrasound system for preclinical investigations
Ben A. Nguyen1  Victor Frenkel1  Pavlos Anastasiadis1  Ali Mohammadabadi1  Meyer J. Fishman1  Jesse A. Smith1  David S. Hersh2 
[1] Department of Diagnostic Radiology and Nuclear Medicine, University of Maryland School of Medicine;
[2] Department of Neurosurgery, University of Maryland School of Medicine;
关键词: Focused ultrasound;    Laser targeting;    Simulations;    Hydrophone measurements;    Hydrogel phantoms;    Acoustic transmission;   
DOI  :  10.1186/s12938-019-0656-z
来源: DOAJ
【 摘 要 】

Abstract Background The clinical applications of transcranial focused ultrasound continue to expand and include ablation as well as drug delivery applications in the brain, where treatments are typically guided by MRI. Although MRI-guided focused ultrasound systems are also preferred for many preclinical investigations, they are expensive to purchase and operate, and require the presence of a nearby imaging center. For many basic mechanistic studies, however, MRI is not required. The purpose of this study was to design, construct, characterize and evaluate a portable, custom, laser-guided focused ultrasound system for noninvasive, transcranial treatments in small rodents. Methods The system comprised an off-the-shelf focused ultrasound transducer and amplifier, with a custom cone fabricated for direct coupling of the transducer to the head region. A laser-guidance apparatus was constructed with a 3D stage for accurate positioning to 1 mm. Pressure field simulations were performed to demonstrate the effects of the coupling cone and the sealing membrane, as well as for determining the location of the focus and acoustic transmission across rat skulls over a range of sizes. Hydrophone measurements and exposures in hydrogels were used to assess the accuracy of the simulations. In vivo treatments were performed in rodents for opening the blood–brain barrier and to assess the performance and accuracy of the system. The effects of varying the acoustic pressure, microbubble dose and animal size were evaluated in terms of efficacy and safety of the treatments. Results The simulation results were validated by the hydrophone measurements and exposures in the hydrogels. The in vivo treatments demonstrated the ability of the system to open the blood–brain barrier. A higher acoustic pressure was required in larger-sized animals, as predicted by the simulations and transmission measurements. In a particular sized animal, the degree of blood–brain barrier opening, and the safety of the treatments were directly associated with the microbubble dose. Conclusion The focused ultrasound system that was developed was found to be a cost-effective alternative to MRI-guided systems as an investigational device that is capable of accurately providing noninvasive, transcranial treatments in rodents.

【 授权许可】

Unknown   

  文献评价指标  
  下载次数:0次 浏览次数:1次