期刊论文详细信息
Química Nova 卷:34
Development and validation of a high performance liquid chromatographic method for determination of cyclosporine-A from biodegradable intraocular implants
Herman Sander Mansur1  Elionai Cassiana de Lima Gomes1  Juliana Barbosa Saliba1  Armando da Silva Cunha Junior1  Gisele Rodrigues da Silva2 
[1] Universidade Federal de Minas Gerais;
[2] Universidade Federal de São João del-Rei;
关键词: cyclosporine-A;    biodegradable intraocular implant;    validation;   
DOI  :  10.1590/S0100-40422011000100026
来源: DOAJ
【 摘 要 】

An HPLC method was developed and validated aiming to quantify the cyclosporine-A incorporated into intraocular implants, released from them; and in direct contact with the degradation products of PLGA. The separation was carried out in isocratic mode using acetonitrile/water (70:30) as mobile phase, a C18 column at 80 ºC and UV detection at 210 nm. The method provided selectivity based on resolution among peaks. It was linear over the range of 2.5-40.0 µg/mL. The quantitation and detection limits were 0.8 and 1.2 µg/mL, respectively. The recovery was 101.8% and intra-day and inter-day precision was close to 2%.

【 授权许可】

Unknown   

  文献评价指标  
  下载次数:0次 浏览次数:0次