Development and validation of a high performance liquid chromatographic method for determination of cyclosporine-A from biodegradable intraocular implants
Juliana Barbosa Saliba2  Armando Da Silva Cunha Junior2  Elionai Cassiana De Lima Gomes2  Herman Sander Mansur1  Gisele Rodrigues Da Silva1 
[1] ,Universidade Federal de Minas Gerais Faculdade de Farmácia Belo Horizonte MG ,Brasil
关键词: cyclosporine-A;    biodegradable intraocular implant;    validation;   
DOI  :  10.1590/S0100-40422011000100026
来源: SciELO
PDF
【 摘 要 】

An HPLC method was developed and validated aiming to quantify the cyclosporine-A incorporated into intraocular implants, released from them; and in direct contact with the degradation products of PLGA. The separation was carried out in isocratic mode using acetonitrile/water (70:30) as mobile phase, a C18 column at 80 ºC and UV detection at 210 nm. The method provided selectivity based on resolution among peaks. It was linear over the range of 2.5-40.0 µg/mL. The quantitation and detection limits were 0.8 and 1.2 µg/mL, respectively. The recovery was 101.8% and intra-day and inter-day precision was close to 2%.