期刊论文详细信息
Journal of Medical Radiation Sciences
Whole breast and excision cavity radiotherapy plan comparison: Conformal radiotherapy with sequential boost versus intensity‐modulated radiation therapy with a simultaneously integrated boost
Katherine Small1  Chris Kelly1  Rachael Beldham-Collins1 
[1] Nepean Cancer Care Centre, Penrith, New South Wales 2750, Australia
关键词: Breast cancer;    conformal radiotherapy;    intensity‐modulated radiotherapy (IMRT);    lung dose;    radiation therapy;   
DOI  :  10.1002/jmrs.4
来源: Wiley
PDF
【 摘 要 】

Abstract

Introduction

A comparative study was conducted comparing the difference between (1) conformal radiotherapy (CRT) to the whole breast with sequential boost excision cavity plans and (2) intensity-modulated radiation therapy (IMRT) to the whole breast with simultaneously integrated boost to the excision cavity. The computed tomography (CT) data sets of 25 breast cancer patients were used and the results analysed to determine if either planning method produced superior plans.

Methods

CT data sets from 25 past breast cancer patients were planned using (1) CRT prescribed to 50 Gy in 25 fractions (Fx) to the whole-breast planning target volume (PTV) and 10 Gy in 5Fx to the excision cavity and (2) IMRT prescribed to 60 Gy in 25Fx, with 60 Gy delivered to the excision cavity PTV and 50 Gy delivered to the whole-breast PTV, treated simultaneously. In total, 50 plans were created, with each plan evaluated by PTV coverage using conformity indices, plan maximum dose, lung dose, and heart maximum dose for patients with left-side lesions.

Results

CRT plans delivered the lowest plan maximum doses in 56% of cases (average CRT = 6314.34 cGy, IMRT = 6371.52 cGy). They also delivered the lowest mean lung dose in 68% of cases (average CRT = 1206.64 cGy, IMRT = 1288.37 cGy) and V20 in 88% of cases (average CRT = 20.03%, IMRT = 21.73%) and V30 doses in 92% of cases (average CRT = 16.82%, IMRT = 17.97%). IMRT created more conformal plans, using both conformity index and conformation number, in every instance, and lower heart maximum doses in 78.6% of cases (average CRT = 5295.26 cGy, IMRT = 5209.87 cGy).

Conclusion

IMRT plans produced superior dose conformity and shorter treatment duration, but a slightly higher planning maximum and increased lung doses. IMRT plans are also faster to treat on a daily basis, with shorter fractionation.

【 授权许可】

CC BY-NC   
© 2013 The Authors. Journal of Medical Radiation Sciences published by Wiley Publishing Asia Pty Ltd on behalf of Australian Institute of Radiography and New Zealand Institute of Medical Radiation Technology.

Creative Commons Attribution-NonCommercial License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited and is not used for commercial purposes.

【 预 览 】
附件列表
Files Size Format View
RO202107150014329ZK.pdf 1908KB PDF download
  文献评价指标  
  下载次数:1次 浏览次数:1次