Background
Despite being used commonly in bovine medicine, information on reliability of point-of-care (POC) lactate meters is limited.
Journal of Veterinary Internal Medicine | |
Evaluation of 4 Point‐of‐Care Units for the Determination of Blood l‐Lactate Concentration in Cattle | |
T. Karapinar2  O. Kaynar1  A. Hayirli3  | |
[1] Faculty of Veterinary Medicine, Department of Biochemistry, Ataturk University, Erzurum, Turkey;Faculty of Veterinary Medicine, Department of Internal Medicine, Firat University, Elazig, Turkey;Faculty of Veterinary Medicine, Department of Animal Nutrition and Nutritional Disorders, Ataturk University, Erzurum, Turkey | |
关键词: Accuracy; Bovine; Portable lactate meter; Precision; Sensitivity; Specificity; | |
DOI : 10.1111/jvim.12204 | |
来源: Wiley | |
Despite being used commonly in bovine medicine, information on reliability of point-of-care (POC) lactate meters is limited. To determine the validity of 4 commercially available POC lactate meters in cattle. Cattle with various diseases (n = 120). Blood samples collected from the jugular vein were processed immediately on POC lactate meters. Plasma l-lactate concentration was measured by the enzymatic-colorimetric method (ELISA). Data were subjected to Friedman's test for comparison, Passing-Bablok regression and Bland-Altman plot analyses for reliability, and receiver operating characteristics analysis for sensitivity (Se) and specificity (Sp). The POC lactate meters were highly correlated with ELISA (r = 0.98–0.99) despite disagreements among units. When regressed on ELISA, blood l-lactate concentrations generated from Accutrend Plus and Lactate Pro were linear up to 16.6 and 15.7 mmol/L, respectively, whereas those generated from i-STAT and Lactate Scout were linear up to 19.5 and 19.7 mmol/L, respectively. All POC lactate meters had a Se of 100% with Sp of 95.7–98.6% at a plasma l-lactate cut-off concentration of 4 mmol/L. i-STAT had the best accuracy (99.0%) and precision (99.8%), the best linear fit (y = −0.13 + 1.04X) yielding the lowest bias (−6.6%) as well as the highest Se (100%) and Sp (98.6%). Despite high correlation with the reference method, dilution is needed for Accutrend Plus/Lactate Pro and i-STAT/Lactate Scout if concentrations >15 and 20 mmol/L, respectively. i-STAT provided the most accurate and precise results.Abstract
Background
Objective
Animals
Methods
Results
Conclusions and Clinical Importance
Unknown
Copyright © 2013 by the American College of Veterinary Internal Medicine
Files | Size | Format | View |
---|---|---|---|
RO202107150007179ZK.pdf | 282KB | download |