Clinical and Experimental Dental Research | |
Accuracy of implant surgery with surgical guide by inexperienced clinicians: an in vitro study | |
Takeshi Toyoshima2  Hideaki Tanaka2  Masanori Sasaki1  Eiji Ichimaru3  Yasushi Naito3  Yasuyuki Matsushita1  Kiyoshi Koyano1  | |
[1] Regenerative Dentistry and Implant Center, Kyushu University Hospital, Fukuoka, Japan;Division of Maxillofacial Diagnostic and Surgical Sciences, Faculty of Dental Science, Kyushu University, Fukuoka, Japan;Ichimaru Dental Office Kuranoue, Shin-Tosu Periodontal and Implant Dentistry Center, Tosu, Japan | |
关键词: artificial mandible; inexperienced clinician; in vitro; implant surgery; model; surgical guide; unilateral free‐end edentulism; | |
DOI : 10.1002/cre2.3 | |
来源: Wiley | |
【 摘 要 】
Implant surgery with surgical guide has been introduced with a concept of position improvement. The surgery might be considered as easy even for inexperienced clinician because of step simplicity. However, there are residual risks, resulting in postoperative complications. The aim of this study was to assess the accuracy of implant surgery with surgical guide by inexperienced clinicians in in vitro. After preoperative computed tomographies (CTs) of five artificial models of unilateral free-end edentulism with scan templates, five surgical guides were established from templates. Following virtual planning, 10 implants were placed in the 45 and 47 regions by five residents without any placement experiences. All drillings and placements were performed using surgical guides. After postoperative CTs, inaccurate verifications between virtual and actual positions of implants were carried out, by overlaying of pre/postoperative CT data. The angle displacement of implant axis in the 47 region was significantly larger than that in the 45 region (P = 0.031). The 3D offset of implant base in the 47 region was significantly larger than that in the 45 region (P = 0.002). For distal/apical directions, displacements of base in the 47 region were significantly larger than those in the 45 region (P = 0.004 and P = 0.003, respectively). The 3D offset of implant tip in the 47 region was significantly larger than that in the 45 region (P = 0.003). For distal/apical directions, displacements of tip in the 47 region were significantly larger than those in the 45 region (P = 0.002 and P = 0.003, respectively). Within limitations of this in vitro study, data for accuracy of implant surgery with surgical guide would be informative for further studies, because in vitro studies should be substantially made to avoid unnecessary burden of patients, in advance of retro/prospective studies. A comparison of the accuracy in this in vitro model between by inexperienced and well-experienced operators should be necessary for clinicians intending to use surgical guide for placement.Abstract
【 授权许可】
CC BY
©2015 The Authors. Clinical and Experimental Dental Research published by John Wiley & Sons Ltd.
Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
RO202107150003278ZK.pdf | 955KB | download |