| Submillimetre thermistors for balloon-borne applications up to lower stratosphere: preliminary characterization with 0.02 K uncertainty | |
| article | |
| F. Sparasci1  D. Jouin1  T. Deuzé1  J. Bordereau2  G. Coeur-Joly2  D. Sourgen2  A. Hertzog2  | |
| [1] Laboratoire Commun de Métrologie LNE-CNAM;Laboratoire de Météorologie Dynamique | |
| 关键词: calibration; thermistors; uncertainty; stratosphere; metrological characterization; | |
| DOI : 10.1002/met.1504 | |
| 学科分类:社会科学、人文和艺术(综合) | |
| 来源: Wiley | |
PDF
|
|
【 摘 要 】
The French metrological laboratory LNE-CNAM, in the framework of a study commissioned by the Laboratory of Dynamic Meteorology LMD, has carried out a preliminary metrological characterization of a device based on submillimetre thermistors, to be used for atmospheric temperature and humidity measurements from the ground up to the lower stratosphere. The device is a measuring system intended to perform temperature measurements between 173 K (around -100 °C) and 233 K (around −40 °C). It is composed of a self-manufactured electronic device, in which submillimetre thermistors measure both air temperature and the temperature of a hygrometer sensor, in order to determine accurately water vapour concentration. Since the needed final measurement uncertainty is below 0.05 K, a comprehensive metrological characterization must be performed. One instrument prototype was supplied to the LNE-CNAM, to be characterized between 173 K and 273 K (around 0 °C). Measurements were carried out in the quasi-adiabatic calorimeter used by the LNE-CNAM to realize the national temperature standard below 84 K. The calorimeter was adapted to work at temperatures up to 273 K. At any setpoint between 173 K and 273 K, the stability of the temperature was within 5 mK for several hours. The thermistors were calibrated by comparison to a calibrated reference thermometer. Preliminary results showed that calibrations with an uncertainty of around 0.02 K are possible. Tests are going to be performed in situ by the LMD, and further characterizations will be carried out to assess the obtained uncertainties and check long-term reproducibility.
【 授权许可】
CC BY|CC BY-NC|CC BY-NC-ND
【 预 览 】
| Files | Size | Format | View |
|---|---|---|---|
| RO202107100001969ZK.pdf | 1012KB |
PDF