Radiation exposure of a mobile 3D C-arm with large flat-panel detector for intraoperative imaging and navigation - an experimental study using an anthropomorphic Alderson phantom
Ulrich Hubbe1  Jan-Helge Klingler1  Christoph Scholz1  Yashar Naseri2  Marie T. Krüger2  Johannes Brönner3 
[1] Department of Neurosurgery, Medical Center - University of Freiburg, Faculty of Medicine, University of Freiburg, Breisacher Str. 64, 79106, Freiburg, Germany;Department of Neurosurgery, Medical Center - University of Freiburg, Faculty of Medicine, University of Freiburg, Breisacher Str. 64, 79106, Freiburg, Germany;Department of Neurosurgery, Cantonal Hospital St. Gallen, St. Gallen, Switzerland;Helmholtz Zentrum München, German Research Center for Environmental Health, Individual Monitoring Service, Munich, Germany;
关键词: 3-dimensional;    C-arm;    Dosimetry;    Intraoperative imaging;    Minimally invasive surgery;    Navigation;    Phantom;    Radiation exposure;    Spine;   
DOI  :  10.1186/s12880-020-00495-y
来源: Springer
PDF
【 摘 要 】

BackgroundIntraoperative 3-dimensional (3D) navigation is increasingly being used for pedicle screw placement. For this purpose, dedicated mobile 3D C-arms are capable of providing intraoperative fluoroscopy-based 3D image data sets. Modern 3D C-arms have a large field of view, which suggests a higher radiation exposure. In this experimental study we therefore investigate the radiation exposure of a new mobile 3D C-arm with large flat-panel detector to a previously reported device with regular flat-panel detector on an Alderson phantom.MethodsWe measured the radiation exposure of the Vision RFD 3D (large 30 × 30 cm detector) while creating 3D image sets as well as standard fluoroscopic images of the cervical and lumbar spine using an Alderson phantom. The dosemeter readings were then compared with the radiation exposure of the previous model Vision FD Vario 3D (smaller 20 × 20 cm detector), which had been examined identically in advance and published elsewhere.ResultsThe larger 3D C-arm induced lower radiation exposures at all dosemeter sites in cervical 3D scans as well as at the sites of eye lenses and thyroid gland in lumbar 3D scans. At ​​male and especially female gonads in lumbar 3D scans, however, the larger 3D C-arm showed higher radiation exposures compared with the smaller 3D C-arm. In lumbar fluoroscopic images, the dosemeters near/in the radiation field measured a higher radiation exposure using the larger 3D C-arm.ConclusionsThe larger 3D C-arm offers the possibility to reduce radiation exposures for specific applications despite its larger flat-panel detector with a larger field of view. However, due to the considerably higher radiation exposure of the larger 3D C-arm during lumbar 3D scans, the smaller 3D C-arm is to be recommended for short-distance instrumentations (mono- and bilevel) from a radiation protection point of view. The larger 3D C-arm with its enlarged 3D image set might be used for long instrumentations of the lumbar spine. From a radiation protection perspective, the use of the respective 3D C-arm should be based on the presented data and the respective application.

【 授权许可】

CC BY   

【 预 览 】
附件列表
Files Size Format View
RO202104246533640ZK.pdf 904KB PDF download
  文献评价指标  
  下载次数:10次 浏览次数:21次