Materials Research | |
Electrochemical behavior of three CP titanium dental implants in artificial saliva | |
Norma Terumi Kadowaki1  Gustavo Aristides Santana Martinez1  Alain Robin1  | |
[1] ,USP EEL LOMLorena SP ,Brazil | |
关键词: dental implants; titanium; corrosion; electrochemical techniques; | |
DOI : 10.1590/S1516-14392009000300018 | |
来源: SciELO | |
【 摘 要 】
The electrochemical behavior of three titanium dental implants purchased on the Brazilian market was evaluated in artificial saliva using open-circuit potential measurements, electrochemical impedance spectroscopy and potentiodynamic polarization. The three implants are passive in this medium but their corrosion resistance and the stability of their passive oxide films are quite different. Although the three implants were made of commercially pure grade titanium, their corrosion current densities and passive current densities were shown to differ from one to two orders of magnitude.
【 授权许可】
CC BY
All the contents of this journal, except where otherwise noted, is licensed under a Creative Commons Attribution License
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
RO202005130152489ZK.pdf | 1318KB | download |