期刊论文详细信息
Journal of the Brazilian Chemical Society
Cathodic voltammetric detection of diltiazem at a bismuth film electrode: application to human urine and pharmaceuticals
Rita I. L. Catarino2  M. Fernanda C. Leal2  Adriana M. Pimenta2  M. Renata S. Souto1  Joana R. T. Lopes2 
[1] ,University Fernando Pessoa Faculty of Health Sciences CEBIMEDPorto,Portugal
关键词: bismuth film electrode;    diltiazem;    voltammetry;    pharmaceuticals;   
DOI  :  10.5935/0103-5053.20140067
来源: SciELO
PDF
【 摘 要 】

In this work it was developed a novel application of bismuth film electrodes (BiFEs) for the determination of the antihypertensive and coronary vasodilator diltiazem by square wave cathodic voltammetry. The bismuth film was deposited ex situ on a glassy carbon electrode for 90 s at -1.4 V vs. Ag/AgCl, from an acetate buffer (pH 4.5; 0.10 mol L-1) containing 5 or 30 mg L-1 Bi. Diltiazem analytical signal was obtained in phosphate buffer (pH 7.4; 0.25 mol L-1) where reduction takes place at -1.5 V vs. Ag/AgCl. The proposed methodology was applied to the quantification of diltiazem in pharmaceutical samples (dynamic linear range comprised between 90 and 900 µg L-1) and in human urine (dynamic linear range comprised between 45 and 270 µg L-1, and detection limit of 12 µg L-1).

【 授权许可】

CC BY   
 All the contents of this journal, except where otherwise noted, is licensed under a Creative Commons Attribution License

【 预 览 】
附件列表
Files Size Format View
RO202005130107347ZK.pdf 281KB PDF download
  文献评价指标  