期刊论文详细信息
Journal of the Brazilian Chemical Society
Evaluation of a continuum source tungsten coil atomic absorption spectrometer: a study of Zn behavior
Joaquim A. Nóbrega2  Jennifer Rust1  Clifton P. Calloway, Jr.1  Bradley T. Jones1 
[1] ,Universidade Federal de São Carlos Departamento de Química São Carlos SP ,Brazil
关键词: atomic absorption spectrometry;    tungsten coil atomizer;    continuum source;    zinc;   
DOI  :  10.1590/S0103-50532005000400023
来源: SciELO
PDF
【 摘 要 】

The performance of a tungsten coil atomizer atomic absorption spectrometer with a D2 lamp continuum source, a Czerny-Turner monochromator, and a charge coupled device detector was evaluated for the determination of Zn. The main attractive characteristics of the instrument are its simplicity and the relative low cost of the atomizer. In spite of a poor limit of detection (40 µg L-1, 3 blank standard deviation/slope), Zn was accurately determined in one water and one urine certified reference standards. The recovery in water was 106% and in urine the obtained result is in the acceptable certified range.

【 授权许可】

CC BY   
 All the contents of this journal, except where otherwise noted, is licensed under a Creative Commons Attribution License

【 预 览 】
附件列表
Files Size Format View
RO202005130104994ZK.pdf 135KB PDF download
  文献评价指标  
  下载次数:7次 浏览次数:2次