Journal of the Brazilian Chemical Society | |
Laboratory validation of a passive sampler for SO2 atmospheric monitoring | |
Lícia P. S. Cruz1  Vânia P. Campos1  Jussival A. P. Novaes1  Tania M. Tavares1  | |
[1] ,Universidade Federal da Bahia Instituto de Química Salvador BA ,Brazil | |
关键词: passive sampling; atmospheric sulfur dioxide; air monitoring; | |
DOI : 10.1590/S0103-50532005000100008 | |
来源: SciELO | |
【 摘 要 】
The performance of a passive sampler was tested under exposure in Teflon chamber. The SO2 trapped in impregnated filter with 1.0 x 10-2 mol L-1 Na2CO3 solution was extracted in ultrasound bath with H2O2 solution for 15 min and determined as SO4(2-) by ion chromatography. The sampler presented good performance except for air in total stillness condition (0 m s-1), for very short exposure periods (0.5 h), for atmosphere with O3 > 150 mug m-3 and for high concentrations of reduced sulfur compounds. The relative humidity of the air (20 - 80%) and the exposure to zero concentration did not affect the sampler performance. It showed stability for one month at 4 ºC or for one week at about 25 ºC. Its precision varied within 3.5 and 23% depending of the SO2 concentration and the accuracy varied from 91 to 100%.
【 授权许可】
CC BY
All the contents of this journal, except where otherwise noted, is licensed under a Creative Commons Attribution License
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
RO202005130104913ZK.pdf | 229KB | download |